Clinical validation of a point-of-care antibody test for COVID-19
This article has been Reviewed by the following groups
Listed in
- Evaluated articles (ScreenIT)
Abstract
The objective of this study was to evaluate the performance of a lateral flow antibody test for COVID-19, approved for use in India. Although many point-of-care antibody tests are available globally, they have been subjected to limited clinical validation. This has led to suboptimal outcomes in the field, where antibody tests play a significant role in tracking the immunity of individuals and communities. In this study an antibody test, ImmunoQuick that recognizes antibodies to the Nucleocapsid and Spike proteins of SARS CoV-2 was tested in 100 symptomatic patients with a positive or negative diagnosis of COVID-19, based on RT-PCR results. The overall sensitivity of the test was found to be 86.1% (95% CI: 76.4% to 92.8%) and specificity 100% (95% confidence interval: 73.5% to 100%). The sensitivity reached a peak of 95.7% with samples taken 17 days after the onset of symptoms. Overall, the sensitivity and specificity of the test are sufficient for assessing seroprevalence.
Article activity feed
-
SciScore for 10.1101/2020.12.16.20248303: (What is this?)
Please note, not all rigor criteria are appropriate for all manuscripts.
Table 1: Rigor
Institutional Review Board Statement IRB: The study was approved by the ethics committee of Smt NHL Municipal Medical College, where the M.P serves on the Faculty.
Consent: Study design: Informed consent was taken from eligible individuals in writing.Randomization not detected. Blinding not detected. Power Analysis not detected. Sex as a biological variable not detected. Table 2: Resources
Software and Algorithms Sentences Resources These samples were then processed using an Enzyme-linked immunosorbent assay (ELISA) kit for total antibodies (COVIDscreen Plus) from Trivitron Healthcare Pvt. Ltd. Trivitron Healthcaresuggested: NoneResults from OddPub: We did not detect open data. We also did not detect open …
SciScore for 10.1101/2020.12.16.20248303: (What is this?)
Please note, not all rigor criteria are appropriate for all manuscripts.
Table 1: Rigor
Institutional Review Board Statement IRB: The study was approved by the ethics committee of Smt NHL Municipal Medical College, where the M.P serves on the Faculty.
Consent: Study design: Informed consent was taken from eligible individuals in writing.Randomization not detected. Blinding not detected. Power Analysis not detected. Sex as a biological variable not detected. Table 2: Resources
Software and Algorithms Sentences Resources These samples were then processed using an Enzyme-linked immunosorbent assay (ELISA) kit for total antibodies (COVIDscreen Plus) from Trivitron Healthcare Pvt. Ltd. Trivitron Healthcaresuggested: NoneResults from OddPub: We did not detect open data. We also did not detect open code. Researchers are encouraged to share open data when possible (see Nature blog).
Results from LimitationRecognizer: An explicit section about the limitations of the techniques employed in this study was not found. We encourage authors to address study limitations.Results from TrialIdentifier: No clinical trial numbers were referenced.
Results from Barzooka: We did not find any issues relating to the usage of bar graphs.
Results from JetFighter: We did not find any issues relating to colormaps.
Results from rtransparent:- Thank you for including a conflict of interest statement. Authors are encouraged to include this statement when submitting to a journal.
- Thank you for including a funding statement. Authors are encouraged to include this statement when submitting to a journal.
- No protocol registration statement was detected.
-