Evaluation of the EUROIMMUN Anti-SARS-CoV-2 ELISA Assay for detection of IgA and IgG antibodies
This article has been Reviewed by the following groups
Discuss this preprint
Start a discussion What are Sciety discussions?Listed in
- Evaluated articles (ScreenIT)
Abstract
As the Coronavirus 2019 (COVID-19) pandemic evolves, the development of immunoassays to help determine exposure and potentially predict immunity has become a pressing priority. In this report we present the performance of the EUROIMMUN enzyme-linked immunosorbent assay (ELISA) for semi-quantitative detection of IgA and IgG antibodies in serum and plasma samples using recombinant S1 domain of the SARS-CoV-2 spike protein as antigen. Specimens from patients, with and without COVID-19 infection, were tested at the University of Chicago Clinical Microbiology and Immunology Laboratory. Of 57 samples from COVID-19 PCR-negative patients, including 28 samples positive for common human coronavirus strains, 53 tested negative and 4 tested positive for IgA (93.0% agreement) while 56 tested negative and 1 tested positive for IgG (98.2% agreement). For COVID-19 PCR-positive patients, 29 of 30 (96.7%) samples collected ≥3 days after positive PCR were positive for IgA, and 28 of 28 samples collected ≥4 days after positive PCR were positive for IgG.
The EUROIMMUN Anti-SARS-CoV-2 ELISA Assay demonstrates excellent sensitivity for detection of IgA and IgG antibodies from samples collected ≥3 days and ≥4 days, respectively, after COVID-19 diagnosis by PCR. This assay did not demonstrate cross reaction in any of the 28 samples from patients with common human coronaviruses, including types HKU1, NL63, CV229E, and OC43.
Article activity feed
-
SciScore for 10.1101/2020.05.11.089862: (What is this?)
Please note, not all rigor criteria are appropriate for all manuscripts.
Table 1: Rigor
Institutional Review Board Statement not detected. Randomization not detected. Blinding not detected. Power Analysis not detected. Sex as a biological variable not detected. Table 2: Resources
Antibodies Sentences Resources The EUROIMMUN Anti-SARS-CoV-2 assay is an enzyme-linked immunosorbent assay (ELISA) that provides semi-quantitative in vitro determination of human antibodies of immunoglobulin classes IgA and IgG against severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) in serum or EDTA plasma (6-7). IgG against severe acute respiratory syndrome coronavirus 2suggested: NoneTo detect the bound antibodies, a second incubation is carried out using an enzyme-labelled antihuman IgA or IgG (enzyme … SciScore for 10.1101/2020.05.11.089862: (What is this?)
Please note, not all rigor criteria are appropriate for all manuscripts.
Table 1: Rigor
Institutional Review Board Statement not detected. Randomization not detected. Blinding not detected. Power Analysis not detected. Sex as a biological variable not detected. Table 2: Resources
Antibodies Sentences Resources The EUROIMMUN Anti-SARS-CoV-2 assay is an enzyme-linked immunosorbent assay (ELISA) that provides semi-quantitative in vitro determination of human antibodies of immunoglobulin classes IgA and IgG against severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) in serum or EDTA plasma (6-7). IgG against severe acute respiratory syndrome coronavirus 2suggested: NoneTo detect the bound antibodies, a second incubation is carried out using an enzyme-labelled antihuman IgA or IgG (enzyme conjugate) catalyzing a color reaction. an enzyme-labelled antihuman IgAsuggested: NoneIgG (enzyme conjugate) catalyzing a colorsuggested: NoneSoftware and Algorithms Sentences Resources All statistical analyses were performed using GraphPad Prism version 8.4.1. GraphPad Prismsuggested: (GraphPad Prism, RRID:SCR_002798)Results from OddPub: We did not detect open data. We also did not detect open code. Researchers are encouraged to share open data when possible (see Nature blog).
Results from LimitationRecognizer: An explicit section about the limitations of the techniques employed in this study was not found. We encourage authors to address study limitations.Results from TrialIdentifier: No clinical trial numbers were referenced.
Results from Barzooka: We did not find any issues relating to the usage of bar graphs.
Results from JetFighter: We did not find any issues relating to colormaps.
Results from rtransparent:- No conflict of interest statement was detected. If there are no conflicts, we encourage authors to explicit state so.
- No funding statement was detected.
- No protocol registration statement was detected.
-