Performing Qualitative Mask Fit Testing Without a Commercial Kit: Fit Testing Which Can Be Performed at Home and at Work
This article has been Reviewed by the following groups
Discuss this preprint
Start a discussion What are Sciety discussions?Listed in
- Evaluated articles (ScreenIT)
Abstract
Objective:
Qualitative fit testing is a popular method of ensuring the fit of sealing face masks such as N95 and FFP3 masks. Increased demand due to the coronavirus disease 2019 (COVID-19) pandemic has led to shortages in testing equipment and has forced many institutions to abandon fit testing. Three key materials are required for qualitative fit testing: the test solution, nebulizer, and testing hood. Accessible alternatives to the testing solution have been studied. This exploratory qualitative study evaluates alternatives to the nebulizer and hoods for performing qualitative fit testing.
Methods:
Four devices were trialed to replace the test kit nebulizer. Two enclosures were tested for their ability to replace the test hood. Three researchers evaluated promising replacements under multiple mask fit conditions to assess functionality and accuracy.
Results:
The aroma diffuser and smaller enclosures allowed participants to perform qualitative fit tests quickly and with high accuracy.
Conclusions:
Aroma diffusers show significant promise in their ability to allow individuals to quickly, easily, and inexpensively perform qualitative fit testing. Our findings indicate that aroma diffusers and homemade testing hoods may allow for qualitative fit testing when conventional apparatus is unavailable. Additional research is needed to evaluate the safety and reliability of these devices.
Article activity feed
-
-
SciScore for 10.1101/2020.08.04.20168344: (What is this?)
Please note, not all rigor criteria are appropriate for all manuscripts.
Table 1: Rigor
NIH rigor criteria are not applicable to paper type.Table 2: Resources
No key resources detected.
Results from OddPub: We did not detect open data. We also did not detect open code. Researchers are encouraged to share open data when possible (see Nature blog).
Results from LimitationRecognizer: An explicit section about the limitations of the techniques employed in this study was not found. We encourage authors to address study limitations.Results from TrialIdentifier: No clinical trial numbers were referenced.
Results from Barzooka: We did not find any issues relating to the usage of bar graphs.
Results from JetFighter: We did not find any issues relating to colormaps.
Results from rtransparent:- Thank…
SciScore for 10.1101/2020.08.04.20168344: (What is this?)
Please note, not all rigor criteria are appropriate for all manuscripts.
Table 1: Rigor
NIH rigor criteria are not applicable to paper type.Table 2: Resources
No key resources detected.
Results from OddPub: We did not detect open data. We also did not detect open code. Researchers are encouraged to share open data when possible (see Nature blog).
Results from LimitationRecognizer: An explicit section about the limitations of the techniques employed in this study was not found. We encourage authors to address study limitations.Results from TrialIdentifier: No clinical trial numbers were referenced.
Results from Barzooka: We did not find any issues relating to the usage of bar graphs.
Results from JetFighter: We did not find any issues relating to colormaps.
Results from rtransparent:- Thank you for including a conflict of interest statement. Authors are encouraged to include this statement when submitting to a journal.
- Thank you for including a funding statement. Authors are encouraged to include this statement when submitting to a journal.
- No protocol registration statement was detected.
-
