Performance of six SARS-CoV-2 immunoassays in comparison with microneutralisation
This article has been Reviewed by the following groups
Listed in
- Evaluated articles (ScreenIT)
Abstract
No abstract available
Article activity feed
-
-
SciScore for 10.1101/2020.05.18.20101618: (What is this?)
Please note, not all rigor criteria are appropriate for all manuscripts.
Table 1: Rigor
Institutional Review Board Statement not detected. Randomization not detected. Blinding not detected. Power Analysis not detected. Sex as a biological variable Serum samples comprising the negative panel: The negative panel consisted of 81 serum samples (from 81 individuals) (median age 64 years, range 2–89 years; 33 males, 31 females). Cell Line Authentication not detected. Table 2: Resources
Antibodies Sentences Resources Thirty-nine out of 81 samples contained autoantibodies: 21 had anti-nuclear antibodies in Hep-2 cell IFA analysis [NOVA Lite® DAPI ANA Kit, Inova Diagnostics, California, USA; staining patterns: homogenous, 7/21; speckled 7/21; centromere 4/21; centromere+anti-mitochondrial antibodies 1/21; … SciScore for 10.1101/2020.05.18.20101618: (What is this?)
Please note, not all rigor criteria are appropriate for all manuscripts.
Table 1: Rigor
Institutional Review Board Statement not detected. Randomization not detected. Blinding not detected. Power Analysis not detected. Sex as a biological variable Serum samples comprising the negative panel: The negative panel consisted of 81 serum samples (from 81 individuals) (median age 64 years, range 2–89 years; 33 males, 31 females). Cell Line Authentication not detected. Table 2: Resources
Antibodies Sentences Resources Thirty-nine out of 81 samples contained autoantibodies: 21 had anti-nuclear antibodies in Hep-2 cell IFA analysis [NOVA Lite® DAPI ANA Kit, Inova Diagnostics, California, USA; staining patterns: homogenous, 7/21; speckled 7/21; centromere 4/21; centromere+anti-mitochondrial antibodies 1/21; nucleolar 1/21; speckled and nuclear dots 1/21], 10 were positive for phospholipase receptor A2 (PLA2R) antibodies (Anti-Phospholipase A2 Receptor IIFT (IgG) (Euroimmun, Lübeck, Germany), three for glomerular basement membrane (GBM) antibodies (EUROPLUS kidney (monkey) and GBM antigen IIFT, Euroimmun, Lübeck, Germany), and five for antineutrophil cytoplasmic (ANCA) antibodies (NOVA Lite® ANCA IFA Kit (Inova Diagnostics, California, USA) (Table 1). anti-nuclearsuggested: NonePLA2Rsuggested: NoneAnti-Phospholipase A2 Receptor IIFT ( IgGsuggested: Noneantineutrophil cytoplasmic ( ANCAsuggested: NoneAutomated immunoassays for anti-SARS-CoV-2 IgG or IgA detection: The analysis of SARS-CoV-2 IgG or IgA antibodies were carried out using the Architect Plus i2000sr Analyzer (Abbott, Illinois, USA) and SARS-COV-2 IgG kit (nucleoprotein based antigen; Abbott; CE marked), EUROLabworkstation (Euroimmun, Lübeck, Germany) and SARS-CoV-2 IgG and IgA kits (S1-based antigen; Euroimmun) and Diasorin Liaison® XL (DiaSorin, Saluggia, Italy) and SARS-CoV-2 S1/S2 IgG kit (S1/S2 based antigen; DiaSorin; research use only) according to manufacturers´ instructions. anti-SARS-CoV-2 IgGsuggested: NoneIgAsuggested: (Lab Vision Cat# RM-9118-S1, RRID:AB_149966)DiaSorin, Saluggia, Italysuggested: NoneExperimental Models: Cell Lines Sentences Resources Thirty-nine out of 81 samples contained autoantibodies: 21 had anti-nuclear antibodies in Hep-2 cell IFA analysis [NOVA Lite® DAPI ANA Kit, Inova Diagnostics, California, USA; staining patterns: homogenous, 7/21; speckled 7/21; centromere 4/21; centromere+anti-mitochondrial antibodies 1/21; nucleolar 1/21; speckled and nuclear dots 1/21], 10 were positive for phospholipase receptor A2 (PLA2R) antibodies (Anti-Phospholipase A2 Receptor IIFT (IgG) (Euroimmun, Lübeck, Germany), three for glomerular basement membrane (GBM) antibodies (EUROPLUS kidney (monkey) and GBM antigen IIFT, Euroimmun, Lübeck, Germany), and five for antineutrophil cytoplasmic (ANCA) antibodies (NOVA Lite® ANCA IFA Kit (Inova Diagnostics, California, USA) (Table 1). Hep-2suggested: NoneBriefly, Vero E6 cells (5⌎×⌎10^4/well) were plated on 96 w/p MEM supplemented with 10% of heat-inactivated FBS, glutamine, penicillin and streptomycin. Vero E6suggested: NoneSoftware and Algorithms Sentences Resources Automated immunoassays for anti-SARS-CoV-2 IgG or IgA detection: The analysis of SARS-CoV-2 IgG or IgA antibodies were carried out using the Architect Plus i2000sr Analyzer (Abbott, Illinois, USA) and SARS-COV-2 IgG kit (nucleoprotein based antigen; Abbott; CE marked), EUROLabworkstation (Euroimmun, Lübeck, Germany) and SARS-CoV-2 IgG and IgA kits (S1-based antigen; Euroimmun) and Diasorin Liaison® XL (DiaSorin, Saluggia, Italy) and SARS-CoV-2 S1/S2 IgG kit (S1/S2 based antigen; DiaSorin; research use only) according to manufacturers´ instructions. Architect Plussuggested: NoneAll samples from the negative panel (N=81) and the patient panel (N=70) were tested with Abbott SARS-CoV-2 IgG, and Euroimmun SARS-CoV-2 IgA and IgG. Abbottsuggested: (Abbott, RRID:SCR_010477)Results from OddPub: We did not detect open data. We also did not detect open code. Researchers are encouraged to share open data when possible (see Nature blog).
Results from LimitationRecognizer: An explicit section about the limitations of the techniques employed in this study was not found. We encourage authors to address study limitations.Results from TrialIdentifier: No clinical trial numbers were referenced.
Results from Barzooka: We did not find any issues relating to the usage of bar graphs.
Results from JetFighter: We did not find any issues relating to colormaps.
Results from rtransparent:- Thank you for including a conflict of interest statement. Authors are encouraged to include this statement when submitting to a journal.
- Thank you for including a funding statement. Authors are encouraged to include this statement when submitting to a journal.
- No protocol registration statement was detected.
-